Disomat Tersus Operating Manual English | Version

| Symptom | Probable Cause | User Action | |---------|---------------|--------------| | Unit won’t power on | Loose DC cord or blown fuse | Check rear fuse holder; replace with T2A 250V fuse | | Rotor won’t spin | Cover not fully closed | Clean magnetic latch contacts with dry cloth | | Erratic OD readings | Condensation on optical window | Run from Maintenance menu (30 sec fan) | | Waste sensor false full | Debris on sensor | Clean with compressed air (max 30 psi) | | Error E07 (rotor imbalance) | Tubes not symmetrically loaded | Remove all tubes, reload with opposing positions filled |

A: Section 8.3 explicitly states that motor replacement requires a certified technician due to polarity and EMC shielding. DIY attempts void warranty. Disomat Tersus Operating Manual English Version